LINEAR 7 — Class II medical device recall Z-2223-2023
Ongoing recall by Datascope Corp., reported 2023-08-02, distributed nationwide. During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introdu
| Recall number | Z-2223-2023 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Datascope Corp., Fairfield, NJ, United States |
| Recall initiated | 2023-06-23 |
| Classified | 2023-07-21 |
| Reported by FDA | 2023-08-02 |
| Distributed | Nationwide (US) |
| Quantity | 321,609 kits |
| Reason classes | device malfunction |
| UPC / GTIN / UDI-DI | 10607567106526, 10607567106533, 10607567106564, 10607567106571, 10607567106632, 10607567106649, 10607567109305, 10607567109312, 10607567109343, 10607567109350, 10607567109480, 10607567109497, 10607567109527, 10607567109534, 10607567109671, 10607567109688, 10607567109718, 10607567109725, 10607567113159, 10607567113197, 10607567113203, 10607567113210 |
Product
LINEAR 7.5Fr.IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES D684-00-0479-01 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (APA) D684-00-0479-02 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (CHINA) D684-00-0479-01C LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (APA)(CHINA) D684-00-0479-02C LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (INDIA) D684-00-0479-07 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (APA) (INDIA) D684-00-0479-08 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (TURKEY) D684-00-0479-09 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES WITH APA (TURKEY) D684-00-0479-10 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES, US ONLY D684-00-0479-01U LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES, (APA) US ONLY D684-00-0479-02U LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES D684-00-0480-01 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (APA) D684-00-0480-02 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (CHINA) D684-00-0480-01C LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (APA)(CHINA) D684-00-0480-02C LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (INDIA) D684-00-0480-07 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (APA)(INDIA) D684-00-0480-08 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES (TURKEY) D684-00-0480-09 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES WITH APA (TURKEY) D684-00-0480-10 LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES, US ONLY D684-00-0480-01U LINEAR 7.5Fr. 40cc IAB WITH ACCESSORIES, (APA) US ONLY D684-00-0480-02U LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES D684-00-0478-01 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (APA) D684-00-0478-02 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (CHINA) D684-00-0478-01C LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (APA) (CHINA) D684-00-0478-02C LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (INDIA) D684-00-0478-07 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (APA) (INDIA) D684-00-0478-08 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (TURKEY) D684-00-0478-09 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES WITH APA (TURKEY) D684-00-0478-10 LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES, US ONLY D684-00-0478-01U LINEAR 7.5Fr. 25cc IAB WITH ACCESSORIES (APA) US ONLY D684-00-0478-02U PACKAGED INSERTION KIT - LINEAR 7.5 Fr. 25 cc IABs D884-00-0019-12 PACKAGED INSERTION KIT - LINEAR 7.5 Fr. 34/40 cc IABs D884-00-0019-13
Reason for recall
During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2223-2023. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.