Data Foundry

Medical device recalls 2012

Access Total T3 Reagent — Class III medical device recall Z-2227-2012

Terminated recall by Beckman Coulter Inc., reported 2012-08-29, distributed nationwide. A recall was initiated because Beckman Coulter has identified a negative bias in test results when comparing e

Recall numberZ-2227-2012
Product typeMedical device
ClassificationClass III
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBeckman Coulter Inc., Brea, CA, United States
Recall initiated2011-10-11
Classified2012-08-20
Reported by FDA2012-08-29
Terminated2013-03-14
DistributedNationwide (US)
Countries outside the USCA

Product

Access Total T3 Reagent, PN: 33830 The Access Total T3 assay is used for the quantitative determination of triiodothyronine (T3) levels in human serum and plasma in the assessment of thyroid function and diagnosis of thyroid disorders when T3 levels are between 0.1 and 8.0 ng/mL.

Reason for recall

A recall was initiated because Beckman Coulter has identified a negative bias in test results when comparing elevated Total T3 patient samples diluted with Sample Diluent A to samples diluted with Access Total T3 Calibrator S0.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2227-2012. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.