Data Foundry

Medical device recalls 2026

Allura system — Class II medical device recall Z-2228-2026

Ongoing recall by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V., reported 2026-05-27, distributed nationwide. The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020,

Recall numberZ-2228-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmPHILIPS MEDICAL SYSTEMS NEDERLAND B.V., Best, Netherlands
Recall initiated2026-04-28
Classified2026-05-21
Reported by FDA2026-05-27
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BD, BE, BG, BH, BO, BR, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DO, DZ, EC, EG, ES, FI, FR, GB, GR, HK, HR, HU, ID, IE, IL, IN, IR, IT, JO, JP, KR, KW, LB, LK, LT, LV, MA, MT, MX, MY, NG, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, SA, SE, SG, SI, SK, TH, TN, TR, TT, TW, UA, UY, VN, ZA
Reason classespackaging
UPC / GTIN / UDI-DI00884838054189, 00884838054196, 00884838054202, 00884838054219, 00884838054226, 00884838054240, 00884838059030, 00884838059047, 00884838059054, 00884838059061, 00884838059085, 00884838059108, 00884838059115, 00884838059122
Serial numbers01H041985, 1000, 1007, 102, 103, 1038, 104, 1044, 1061, 109, 1090, 1094, 1106, 118, 1187, 1192, 1197, 122, 1223, 123, 1232, 124, 1242, 125, 1259 and 339 more
Model numbers722001, 722002, 722003, 722005, 722006, 722008, 722010, 722011, 722012, 722013, 722023, 722025, 722026, 722027, 722028, 722029, 722035, 722038, 722058, 722059

Product

Allura system; System Code Description (Model Numbers): Allura Xper FD10C (722001), Allura Xper FD10F (722002), Allura Xper FD10 (722003, 722010, 722026), Allura Xper FD10/10 (722005, 722011, 722027), Allura Xper FD20 (722006, 722012, 722028), Allura Xper FD20 Biplane (722008, 722013), Allura Xper FD20 OR Table (722015, 722023, 722035), Allura Xper FD20 Biplane OR Table (722025), Allura Xper FD20/10 (722029), Allura Xper FD20/15 (722058), Allura Xper FD20/15 OR Table (722059), Allura Xper FD20/20 (722038), Allura Xper FD20/20 OR Table (722039), Cardio Vascular-Allura (722400);

Reason for recall

The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2228-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.