Data Foundry

Medical device recalls 2025

RAPIDPoint 500 Systems Wash/Waste Cartridge — Class II medical device recall Z-2230-2025

Ongoing recall by Siemens Healthcare Diagnostics Inc, reported 2025-08-06, distributed nationwide. Siemens Healthcare Diagnostics is recalling the RAPIDPoint 500 Systems Wash/Waste Cartridge Lot WW/08925 due t

Recall numberZ-2230-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics Inc, Norwood, MA, United States
Recall initiated2025-06-26
Classified2025-07-31
Reported by FDA2025-08-06
DistributedNationwide (US)
Countries outside the USCA, CL, JP, PY
Quantity1,477 units
UPC / GTIN / UDI-DI00630414482163
Lot numbersWW/08925
Expiration dates2025-12-03

Product

RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.

Reason for recall

Siemens Healthcare Diagnostics is recalling the RAPIDPoint 500 Systems Wash/Waste Cartridge Lot WW/08925 due to a manufacturing error causing positive and negative bias in Ionized Calcium, Potassium, pH, Chloride, pCO2, Glucose and Lactate analytes and their derivatives. Four complaints were filed related to this event that led to Siemens Healthcare Diagnostics identifying an error in the manufacturing process and initiated corrective actions. Use of the affected device may result in errors during QC which could lead to a delay in patient treatment. If bias generated from this issue is not caught during QC erroneous results may cause temporary or medically reversible adverse health consequences to patients.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2230-2025. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.