Data Foundry

Medical device recalls 2016

2 — Class II medical device recall Z-2238-2016

Terminated recall by Stryker Instruments Div. of Stryker Corporation, reported 2016-07-27, distributed nationwide. A variation in flute depth on the routers was observed. The variation in flute depth may cause the routers to

Recall numberZ-2238-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmStryker Instruments Div. of Stryker Corporation, Portage, MI, United States
Recall initiated2016-06-24
Classified2016-07-19
Reported by FDA2016-07-27
Terminated2017-03-23
DistributedNationwide (US)
Countries outside the USCA, CH, GB, JP, NL
Quantity3,163
Lot numbers15077027, 15077067, 15077097, 15078087, 15105017, 15108017, 15108047, 15108077, 15108087, 15108097, 15146017, 15146027, 15146037, 15146047

Product

2.3MM Tapered Router, Product Number 5407-FA2-023 A router is a cutting accessory used in the following medical applications: Neuro; Spine; Ear, Nose, and Throat (ENT)/Otology /Neurotology/ Otorhinolaryngology; Craniofacial (bones of the skull and supraorbital region); and Sternotomy. They are intended to be used with the Stryker CORE" system.

Reason for recall

A variation in flute depth on the routers was observed. The variation in flute depth may cause the routers to be more susceptible to breakages. Risk to Health: Injury to critical soft tissue requiring surgical intervention may occur.