Data Foundry

Medical device recalls 2012

892 — Class II medical device recall Z-2240-2012

Terminated recall by GE OEC Medical Systems, Inc, reported 2012-08-29, distributed nationwide. GE OEC is recalling certain 9800 Image Intensified Flouroscopic x-ray systems due to customer complaint analys

Recall numberZ-2240-2012
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmGE OEC Medical Systems, Inc, Salt Lake City, UT, United States
Recall initiated2009-11-06
Classified2012-08-22
Reported by FDA2012-08-29
Terminated2013-01-28
DistributedNationwide (US)
Quantity3,595 units
Model numbers5304649, HDS721616PLAT80

Product

892.1650 Image-intensified fluoroscopic x-ray system; the device is designed to provide fluoroscopic and spot film imaging of the patient during diagnostic surgical and interventional procedures. The systems include features specifically designed for use in diagnostic and interventional cardiac imaging procedures, and are also intended for cholangiography, endoscopic, urologic, orthopedic, neurologic, vascular, critical care and emergency room procedures. The systems may be used for other imaging applications at the physician's discretion.

Reason for recall

GE OEC is recalling certain 9800 Image Intensified Flouroscopic x-ray systems due to customer complaint analysis showing that a replacement Hitachi 160 GB Hard Drive, when installed on certain models is susceptible of causing unanticipated system shut downs, no boots, data loss or data mix, or unexpected hard drive corruption.