Data Foundry

Medical device recalls 2020

Beckman Coulter ACCESS Immunoassay Systems — Class II medical device recall Z-2240-2020

Terminated recall by Beckman Coulter Inc., reported 2020-06-10, distributed in AK, AL, AR, AS, AZ, CA, CO, CT…. The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all bioti

Recall numberZ-2240-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBeckman Coulter Inc., Brea, CA, United States
Recall initiated2019-04-30
Classified2020-06-02
Reported by FDA2020-06-10
Terminated2021-12-28
DistributedAK, AL, AR, AS, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY
Countries outside the USAE, AT, AU, BD, BE, BG, BH, BR, CA, CH, CL, CN, CR, CW, CY, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GH, GR, GT, GY, HK, HR, HU, ID, IE, IL, IN, IT, JO, JP, KE, KR, KW, LB, LK, LT, MA, MX, MY, NG, NL, NP, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, RU, SE, SG, SK, SV, TH, TN, TR, TW, UY, VE, VN, ZA
Quantity95,453 units
UPC / GTIN / UDI-DI15099590227173
Model numbers33860

Product

Beckman Coulter ACCESS Immunoassay Systems, Access Thyroglobulin, REF 33860 (UDI: 15099590227173) for use with the Access Family of Immunoassay Systems including: Access 2, UniCel DxI 600, UniCel DxI 800, UniCel DxC 600i, UniCel DxC 660i, UniCel DxC 680i, UniCel DxC 860i, and UniCel DxC 880i systems.

Reason for recall

The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays (Free T3-REF A13422, Total T3-REF 33830, Free T4-REF 33880, GI Mon-387687, Tg-REF 33860, TgAbII-REF A32898 displayed significant interference (defined as a result bias greater than +/-10%) at 1,200 ng/mL of biotin.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2240-2020. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.