Data Foundry

Medical device recalls 2012

EasyLink Informatic System is included with the Dimension Vista System — Class II medical device recall Z-2242-2012

Terminated recall by Siemens Healthcare Diagnostics, Inc., reported 2012-08-29, distributed nationwide. When using the EasyLink Quality Control Package the system may not perform as intended for Multi-Rule QC viola

Recall numberZ-2242-2012
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics, Inc., Newark, DE, United States
Recall initiated2012-05-29
Classified2012-08-22
Reported by FDA2012-08-29
Terminated2014-04-10
DistributedNationwide (US)
Countries outside the USAT, AU, BE, CA, CH, CZ, DE, DK, ES, FR, IT, JP, KR, MY, NL, NO, NZ, PH, PT, SA, SG, SI, SK, TW
Quantity637
Model numbers1000034805

Product

EasyLink Informatic System is included with the Dimension Vista System. The Dimension Vista Integrated System is an in vitro diagnostic device intended tot duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to determine a variety of analytes in human body fluids. The EasyLink Informatics System is classified as a data calculator and processing module for use with in vitro diagnostic devices. The EasyLink is included with the Dimension Vista System and may be used with other Siemens analyzers as a communications and connectivity workstation for integration with laboratory information system (LIS) networks.

Reason for recall

When using the EasyLink Quality Control Package the system may not perform as intended for Multi-Rule QC violations [e.g., 2(2S), 4(1s) and 10x], unless configured using the instructions send with the Urgent Field Safety Notice. When this situation occurs, QC status may not be flagged appropriately. If auto-verification rules are defined to hold patient results when QC status is out, the patient

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2242-2012. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.