Data Foundry

Medical device recalls 2024

VITROS Chemistry Products Calibrator Kit 29- In-Vitor used to… — Class III medical device recall Z-2245-2024

Ongoing recall by Ortho-Clinical Diagnostics, Inc., reported 2024-07-10, distributed nationwide. Kit assigned an extended expiration date of 24-Jul-2025 during release. The correct expiration date for Lot 29

Recall numberZ-2245-2024
Product typeMedical device
ClassificationClass III
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmOrtho-Clinical Diagnostics, Inc., Rochester, NY, United States
Recall initiated2024-05-14
Classified2024-07-01
Reported by FDA2024-07-10
DistributedNationwide (US)
Countries outside the USAU, BE, BM, BR, CA, CL, CN, CO, DE, DK, ES, FR, GB, IN, IT, JP, MX, NL, NO, PT, RU, SE, SG
Quantity1,718
Reason classespotency
UPC / GTIN / UDI-DI10758750002672
Lot numbers2953
Model numbers6802344
Expiration dates2024-07-24

Product

VITROS Chemistry Products Calibrator Kit 29- In-Vitor used to calibrate the VITROS 5,1 FS/4600 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems for the quantitative measurement of total iron-binding capacity (TIBC) using VITROS Chemistry Products dTIBC Reagent Model/Catalog Number: 6802344

Reason for recall

Kit assigned an extended expiration date of 24-Jul-2025 during release. The correct expiration date for Lot 2953 is 24-Jul-2024. If used past its true expiration and the customer does not have their Assay Data Disk (ADD) updated, the system will allow them to use the expired calibrator which may bias the results. However, at this time the calibrator is within its use date and will function appropriately.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2245-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.