Pivot Bipolar Cup 22x42/43-51/53mm — Class II medical device recall Z-2253-2014
Terminated recall by Ortho Development Corporation, reported 2014-08-27, distributed nationwide. Ortho Development is recalling various Pivot Bipolar Cup, Patella, PS Tibial Insert, UC Tibila Insert, and CK
| Recall number | Z-2253-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Ortho Development Corporation, Draper, UT, United States |
| Recall initiated | 2014-07-11 |
| Classified | 2014-08-17 |
| Reported by FDA | 2014-08-27 |
| Terminated | 2014-10-06 |
| Distributed | Nationwide (US) |
| Countries outside the US | JP |
| Quantity | 188 |
| Reason classes | packaging |
| Lot numbers | A136254, A136255, A136256, A136257, A136258, A136259, A136260, A136261, A136262, A136263, A136265, A136266, A136267, A136268, A136269, A136270 |
| Model numbers | 133-2242, 133-2243, 133-2244, 133-2245, 133-2246, 133-2247, 133-2248, 133-2249, 133-2250, 133-2251, 133-2253, 133-YYYY |
Product
Pivot Bipolar Cup 22x42/43-51/53mm, REF 133-YYYY Product Usage: Used in a hemiarthroplasty procedure as the bearing surface to articulate with both the CoCr femoral head and the natural acetabulum
Reason for recall
Ortho Development is recalling various Pivot Bipolar Cup, Patella, PS Tibial Insert, UC Tibila Insert, and CK Tibil Insert because the foil pouches did not properly seal, allowing air to seep back into the pouch after sealing which increases the risk that the material properties of the implant would be substantially degraded due to oxidation.