Data Foundry

Medical device recalls 2014

PS Tibial Insert — Class II medical device recall Z-2255-2014

Terminated recall by Ortho Development Corporation, reported 2014-08-27, distributed nationwide. Ortho Development is recalling various Pivot Bipolar Cup, Patella, PS Tibial Insert, UC Tibila Insert, and CK

Recall numberZ-2255-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmOrtho Development Corporation, Draper, UT, United States
Recall initiated2014-07-11
Classified2014-08-17
Reported by FDA2014-08-27
Terminated2014-10-06
DistributedNationwide (US)
Countries outside the USJP
Quantity11
Reason classespackaging
Lot numbersA135827, A135943, A135945, A136453, A136454, A136492
Model numbers163-1308, 163-1309, 163-1312, 163-1407, 163-1509, 163-1XXX

Product

PS Tibial Insert, REF 163-1XXX Product Usage: UHMWPE component used in TKA procedures to serve as a tibial bearing surface for articulation with the femoral component

Reason for recall

Ortho Development is recalling various Pivot Bipolar Cup, Patella, PS Tibial Insert, UC Tibila Insert, and CK Tibil Insert because the foil pouches did not properly seal, allowing air to seep back into the pouch after sealing which increases the risk that the material properties of the implant would be substantially degraded due to oxidation.