Data Foundry

Medical device recalls 2018

VNS Therapy Programmer — Class II medical device recall Z-2255-2018

Terminated recall by Cyberonics, Inc, reported 2018-06-27, distributed in AL, CA, CO, DE, FL, GA, ID, IL…. Certain Model 3000 programming events can result in miscalculation of parameters stored in the Models 103, 104

Recall numberZ-2255-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmCyberonics, Inc, Houston, TX, United States
Recall initiated2017-11-13
Classified2018-06-19
Reported by FDA2018-06-27
Terminated2022-10-31
DistributedAL, CA, CO, DE, FL, GA, ID, IL, IN, ME, MO, MS, NC, NJ, NY, PA, TN, TX, UT, WA, WI
Quantity252
Reason classessoftware
Serial numbers11R57F2, 1CQ57F2, 1G7G1F2, 1HT77F2, 1JR57F2, 1MYB7F2, 1TQ57F2, 1YR77F2, 2CP57F2, 2DR77F2, 2DT77F2, 2LYB7F2, 4GR57F2, 4GT77F2, 4JS77F2, 4LR77F2, 55T77F2, 57P57F2, 5D0C7F2, 5HS77F2, 5KS77F2, 5MR77F2, 5NQ57F2, 5RQ57F2, 5VQ57F2 and 226 more
Model numbers3000, V1.0

Product

VNS Therapy Programmer, Model 3000, v1.0 System

Reason for recall

Certain Model 3000 programming events can result in miscalculation of parameters stored in the Models 103, 104, 105, and 106 generators. During these programming events, the miscalculations can lead to: " Delivery of more stimulation than intended, resulting in painful stimulation or other common side effects (Model 106 only); " No stimulation in the case of device disablement (Burst Watchdog Timeout), resulting in no therapy to the patient (Model 106 only); " Delivery of less stimulation than intended, resulting in therapeutic settings not being achieved within device specification (Models 103, 104, 105, or 106); and/or " Delays or absence of the 75% and 50% battery life indicators displayed by the programming software (Models 103, 104, 105, or 106).