Azurion system — Class II medical device recall Z-2258-2026
Ongoing recall by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V., reported 2026-06-03, distributed nationwide. It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance a
| Recall number | Z-2258-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V., Best, Netherlands |
| Recall initiated | 2026-04-30 |
| Classified | 2026-05-28 |
| Reported by FDA | 2026-06-03 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AF, AR, AT, AU, BD, BE, BG, BH, BO, BR, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GR, HK, HN, HR, HU, ID, IE, IL, IN, IQ, IR, IT, JO, JP, KE, KR, KW, LB, LK, LT, LV, MA, MT, MX, MY, NG, NI, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, SA, SE, SG, SI, SK, TH, TN, TR, TT, TW, UA, UY, VN, ZA |
| UPC / GTIN / UDI-DI | 00884838085251, 00884838085268, 00884838085275, 00884838085282, 00884838085350, 00884838085367, 00884838099203, 00884838099210, 00884838099227, 00884838099234, 00884838099241, 00884838099258, 00884838099265, 00884838099272 |
| Model numbers | 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222, 722223, 722224, 722225, 722226, 722227, 722228 |
Product
Azurion system; System Code Description (Model Numbers): Azurion 3 M12 (722063, 722221), Azurion 3 M15 (722064, 722222), Azurion 5 M12 (722227), Azurion 5 M20 (722228), Azurion 7 B12 (722067, 722225), Azurion 7 B20 (722068, 722226), Azurion 7 M12 (722078, 722223), Azurion 7 M20 (722079, 722224).
Reason for recall
It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2258-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.