Data Foundry

Medical device recalls 2023

Incisive CT for Brazil SKD-Computed Tomography X-Ray System Model:… — Class II medical device recall Z-2259-2023

Ongoing recall by Philips North America, reported 2023-08-09, distributed nationwide. Metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the Incisive CT system

Recall numberZ-2259-2023
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmPhilips North America, Cambridge, MA, United States
Recall initiated2023-04-25
Classified2023-07-28
Reported by FDA2023-08-09
DistributedNationwide (US)
Countries outside the USAR, AT, AU, BD, BE, BG, BH, BO, BR, CA, CL, CN, CO, CZ, DE, DZ, EC, EG, ES, FR, GB, GR, ID, IL, IN, IQ, IR, IT, JO, JP, KR, LV, MX, NG, NL, PA, PH, PK, PL, RO, RS, SA, SE, SK, TH, TN, TR, TW, VN
Quantity20 units
Reason classesdevice malfunction
Serial numbers540000, 540001, 540002, 540003, 540004, 540005, 540006, 540007, 540008, 540009, 540010, 540011, 540012, 540013, 540014, 540015, 540016, 540017, 540018
Model numbers728146

Product

Incisive CT for Brazil SKD-Computed Tomography X-Ray System Model: 728146

Reason for recall

Metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the Incisive CT system may become expelled and make contact with other components and cause injury to user or if detached a loud noise will be emitted and the system will shut down

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2259-2023. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.