Container fill and transfer sets — Class II medical device recall Z-2260-2016
Terminated recall by The Metrix Company, reported 2016-08-03, distributed in AL, AZ, CA, CO, CT, FL, IA, IL…. An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the design
| Recall number | Z-2260-2016 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | The Metrix Company, Dubuque, IA, United States |
| Recall initiated | 2016-06-23 |
| Classified | 2016-07-22 |
| Reported by FDA | 2016-08-03 |
| Terminated | 2023-04-11 |
| Distributed | AL, AZ, CA, CO, CT, FL, IA, IL, KS, LA, MA, MD, MN, NC, OH, PA, PR, TN, TX, UT, WA |
| Countries outside the US | AU, CA, DO, GB, MY, TR |
| Quantity | 137,598 cases |
Product
Container fill and transfer sets, 250 mL, 500 mL, 1000 mL, 2000 mL, 3000 mL, or 4000 mL.
Reason for recall
An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when dispensing into bags. After reviewing the initial design of the device, it was determined the dispensing volume validation was not properly done and none of the compounders on the market can meet the stated specifications.