Stryker SmartPump Tourniquet — Class II medical device recall Z-2277-2025

Ongoing recall by Stryker Corporation, reported 2025-08-13, distributed nationwide. Tourniquet cuff flange may become detached from the bladder. If leak occurs during the procedure, there may be

Recall numberZ-2277-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmStryker Corporation, Portage, MI, United States
Recall initiated2025-07-02
Classified2025-08-07
Reported by FDA2025-08-13
DistributedNationwide (US)
Countries outside the USAU, CA, GB, HK, IN, MY, NL, NZ, SE, SG, TH
Quantity233.23
Reason classespackaging, device malfunction
UPC / GTIN / UDI-DI07613154599257, 07613154599295, 07613154599370

Product

Stryker SmartPump Tourniquet, disposable sterile (single use) DISP 18X3,1BLA,1PRT QUICK / 5921-018-135; DISP 18X3,1BLA,2PRT QUICK / 5921-018-235; DISP 24X4,1BLA, 2PRT QUICK / 5921-024-235

Reason for recall

Tourniquet cuff flange may become detached from the bladder. If leak occurs during the procedure, there may be risk of hemorrhage, resulting in blood loss greater than expected. Other serious risks include cerebral hypotension and cardiovascular collapse and arrest, risking stroke or death. Specific to IVRA, patient toxicity may also occur due to rapid systemic absorption of anesthetics.