Data Foundry

Medical device recalls 2013

SmartSite Low Sorbing Infusion Set — Class II medical device recall Z-2279-2013

Terminated recall by CareFusion 303, Inc., reported 2013-10-02, distributed in AZ, CA, FL, TN. CareFusion is recalling the SmartSite Low Sorbing Infusion Set, Model 72313E, Lot Number 13016314, because of

Recall numberZ-2279-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmCareFusion 303, Inc., San Diego, CA, United States
Recall initiated2013-08-30
Classified2013-09-25
Reported by FDA2013-10-02
Terminated2013-10-28
DistributedAZ, CA, FL, TN
Quantity600 units
Lot numbers13016314
Model numbers72313E

Product

SmartSite Low Sorbing Infusion Set, Model#72313E The SmartSite Low Sorbing Infusion Sets are used to administer fluid and medications through a needle or catheter inserted into the patient's artery or vein. The SmartSite Low Sorbing set is comprised of components commonly found on intravascular administration sets and extension sets. It includes a 1.2 micron filter, needleless SmartSite connector, low sorbing tubing segment, pinch clamps, and luer connectors.

Reason for recall

CareFusion is recalling the SmartSite Low Sorbing Infusion Set, Model 72313E, Lot Number 13016314, because of an incorrect filter was assembled on this lot. The correct filter is a 1.2 micron filter (blue color) but a 0.2 micron filter (clear color) may have been added instead.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2279-2013. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.