Data Foundry

Medical device recalls 2020

ADVIA Chemistry Fructosamine Reagent — Class II medical device recall Z-2280-2020

Terminated recall by Siemens Healthcare Diagnostics, Inc., reported 2020-06-17, distributed nationwide. The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference i

Recall numberZ-2280-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics, Inc., Newark, DE, United States
Recall initiated2020-05-04
Classified2020-06-05
Reported by FDA2020-06-17
Terminated2021-09-20
DistributedNationwide (US)
Quantity5
UPC / GTIN / UDI-DI00630414558073
Lot numbers501874

Product

ADVIA Chemistry Fructosamine Reagent (FRUC), Siemens Material Number 10361941, UDI Number 00630414558073 - Product Usage: use in the quantitative measurement of glycated protein (fructosamine) in human serum and plasma for monitoring the blood glucose levels of diabetic patients over the preceding two-three weeks.

Reason for recall

The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.