Dimension Enzymatic Creatinine — Class II medical device recall Z-2284-2020
Terminated recall by Siemens Healthcare Diagnostics, Inc., reported 2020-06-17, distributed nationwide. The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference i
| Recall number | Z-2284-2020 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Siemens Healthcare Diagnostics, Inc., Newark, DE, United States |
| Recall initiated | 2020-05-04 |
| Classified | 2020-06-05 |
| Reported by FDA | 2020-06-17 |
| Terminated | 2021-09-20 |
| Distributed | Nationwide (US) |
| Quantity | 8,261 |
| UPC / GTIN / UDI-DI | 00842768030512 |
| Lot numbers | BB0262, EA1102, FB0122, FB1050, FC0324, GC0184, GD0218, GF1006 |
Product
Dimension Enzymatic Creatinine, Siemens Material Number 10471520, UDI Number 00842768030512 - Product Usage: use in the quantitative determination of creatinine in human serum, plasma, and urine for the diagnosis and treatment of renal disease.
Reason for recall
The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.