Data Foundry

Medical device recalls 2018

ABX Pentra Micro ALBUMIN 2 CP — Class II medical device recall Z-2286-2018

Terminated recall by Horiba Instruments, Inc dba Horiba Medical, reported 2018-07-04, distributed nationwide. When the ABX Pentra Micro ALBUMIN 2 CP (ALBT2) Reagent 2 is sampled and dispensed immediately after an ABX Pen

Recall numberZ-2286-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmHoriba Instruments, Inc dba Horiba Medical, Irvine, CA, United States
Recall initiated2018-04-26
Classified2018-06-25
Reported by FDA2018-07-04
Terminated2019-09-26
DistributedNationwide (US)
Countries outside the USBG, DE, FR, IT, LB, PL, PT, RO, SA, TH, ZA
Model numbers1300032563

Product

ABX Pentra Micro ALBUMIN 2 CP, Serum/Plasma, REF 1300032563 The device is intended for use in conjunction with certain materials to measure a variety of analytes.

Reason for recall

When the ABX Pentra Micro ALBUMIN 2 CP (ALBT2) Reagent 2 is sampled and dispensed immediately after an ABX Pentra Creatinine 120 CP (CREA3) Reagent 2 dispense, the final result of the microalbumin may be elevated more than 10%. This increase may affect serum, plasma and urine sample types.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2286-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.