Trilogy EV300 — Class II medical device recall Z-2289-2021
Ongoing recall by Philips Respironics, Inc., reported 2021-08-25, distributed nationwide. Two software issues have been identified related to pressure increase in the device: The first issue is descri
| Recall number | Z-2289-2021 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Philips Respironics, Inc., Murrysville, PA, United States |
| Recall initiated | 2021-07-26 |
| Classified | 2021-08-13 |
| Reported by FDA | 2021-08-25 |
| Distributed | Nationwide (US) |
| Reason classes | software |
Product
Trilogy EV300, Product numbers CA2200X12B, DS2200X11B, FX2200X15B, IN2200X15B
Reason for recall
Two software issues have been identified related to pressure increase in the device: The first issue is described as "Infant/Pediatric EFS Calibration Pressure Increase" and the second issue is described as "Pressure Drift (Continuous Usage).