AU5800 Chemistry Analyzers — Class II medical device recall Z-2310-2016
Terminated recall by Beckman Coulter Inc., reported 2016-08-10, distributed nationwide. Beckman coulter has received complaints regarding leaking and cracked C cups. When the cup cracks, coolant can
| Recall number | Z-2310-2016 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Beckman Coulter Inc., Brea, CA, United States |
| Recall initiated | 2015-07-30 |
| Classified | 2016-08-01 |
| Reported by FDA | 2016-08-10 |
| Terminated | 2017-02-15 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AT, AU, BD, BE, BG, BO, BR, CA, CH, CN, CR, CY, CZ, DE, DK, DZ, EC, EG, ES, FI, FR, GB, GH, GR, HK, HR, HU, ID, IE, IL, IN, IT, JO, JP, KE, KR, KW, LB, LK, LT, MA, MX, MY, NG, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, QA, RO, RS, SA, SE, SG, SI, SK, SV, TH, TN, TR, TW, UA, UY, VN, ZA |
| Reason classes | packaging |
Product
AU5800 Chemistry Analyzers, Reference No. A98352 The AU series are automated, clinical chemistry analyzers that measure analytes in patient samples (serum, plasma, urine, CSF) in combination with appropriate reagents, calibrators, quality, control (QC) materials, and other accessories.
Reason for recall
Beckman coulter has received complaints regarding leaking and cracked C cups. When the cup cracks, coolant can leak onto the connectors of the Printed Circuit Boards (PCB) in the analyzer, resulting in a short-circuit causing the instrument to go into a stop mode.