Data Foundry

Medical device recalls 2021

Cell Marque Antibodies — Class III medical device recall Z-2311-2021

Ongoing recall by Cell Marque Corporation, reported 2021-08-25, distributed nationwide. Mouse Monoclonal Antibody labelled with an incorrect expiry date of 31 Dec 2023. The correct expiry date is 31

Recall numberZ-2311-2021
Product typeMedical device
ClassificationClass III
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmCell Marque Corporation, Rocklin, CA, United States
Recall initiated2021-07-15
Classified2021-08-18
Reported by FDA2021-08-25
DistributedNationwide (US)
Countries outside the USAR, BR, CA, KR, MX, TH
Quantity20
Reason classeslabeling
Lot numbers324M-15/0000116695, 324M-16/0000116696
Model numbers324M-15, 324M-16

Product

Cell Marque Antibodies, PSA (ER-PR8) Mouse Monoclonal Antibody, REF: 324M-15, 324M-16

Reason for recall

Mouse Monoclonal Antibody labelled with an incorrect expiry date of 31 Dec 2023. The correct expiry date is 31 Dec 2022. If used beyond expiry, the material performance is likely to decrease and may be observed as weak or no staining, and there is a potential risk that a false negative result could be recorded and may lead to delay of diagnosis due to retest.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2311-2021. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.