Data Foundry

Medical device recalls 2021

728327 Ingenuity CT Upgrades-Computed Tomography X-ray system — Class II medical device recall Z-2316-2021

Terminated recall by Philips North America Llc, reported 2021-08-25, distributed nationwide. When setting patient weight unit preferences to pounds, SynchRight P3T (Personalized Patient Protocol) softwar

Recall numberZ-2316-2021
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmPhilips North America Llc, Cambridge, MA, United States
Recall initiated2021-06-08
Classified2021-08-18
Reported by FDA2021-08-25
Terminated2023-08-10
DistributedNationwide (US)
Quantity3 units
Reason classessoftware
Serial numbers300099, 300106, 300143

Product

728327 Ingenuity CT Upgrades-Computed Tomography X-ray system

Reason for recall

When setting patient weight unit preferences to pounds, SynchRight P3T (Personalized Patient Protocol) software will result in increased contrast volume, resulting in increased contrast volume recommended for the patient.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2316-2021. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.