Data Foundry

Medical device recalls 2018

Fortify Assura — Class II medical device recall Z-2321-2018

Terminated recall by St Jude Medical Inc., reported 2018-07-04, distributed nationwide. The firm is releasing new firmware for high voltage implantable cardiac devices (ICD) and cardiac resynchroniz

Recall numberZ-2321-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSt Jude Medical Inc., Sylmar, CA, United States
Recall initiated2018-04-16
Classified2018-06-28
Reported by FDA2018-07-04
Terminated2024-03-08
DistributedNationwide (US)
Quantity131,093
Reason classessoftware
Serial numbers1003754, 1003755, 1003756, 1003757, 1003758, 1003759, 1003760, 1003761, 1003762, 1003764, 1003765, 1003766, 1003767, 1003770, 1003772, 1003776, 1003777, 1003778, 1003780, 1003781, 1003782, 1003783, 1003784, 1003785, 1003788 and 475 more
Model numbers100059623, CD1257-40/100042322

Product

Fortify Assura, Sterile EO, Model #/ Part #: CD1257-40/100042322, 100059623; CD1257-40Q/100042341, 100059535; CD1259-40/100041914, 100042338, 100059536, 100059646, 100081567; CD1259-40Q/100041931, 100042339, 100059537, 100059620, 100059626, 100078685; CD1263-40/100042948; CD1263-40Q/100042949; CD1357-40C/100080365, 100080377, 100083441; CD1357-40Q/100080389, 100080390, 100083388; CD1359-40/100078740, 100083267, 100096839; CD1359-40C/100078500, 100078969, 100079372, 100079373, 100083442, 100083443, 100117545; CD1359-40Q/100078804, 100079155, 100079413, 100083445, 100096884; CD1359-40QC/100078603, 100079145, 100079146, 100079239, 100083405, 100083444, 100117632, 100127000, 100127065; CD1363-40C/100080349; CD1363-40Q/100080317; CD1391-40C/100080756; CD1391-40QC/100080783; CD2257-40/100042337, 100059839; CD2257-40Q/100042342, 100059842; CD2259-40/100041893, 100042308, 100059794, 100059907, 100082567; CD2259-40Q/100041894, 100042309, 100059845, 100059873, 100078610; CD2263-40/100042950; CD2263-40Q/100043201; CD2357-40C/100080404, 100080413, 100083490; CD2357-40Q/100080319, 100080412, 100083492; CD2359-40/100078750, 100083466; CD2359-40C/100078651, 100079201, 100079240, 100079442, 100083501, 100083502, 100117546; CD2359-40Q/100078841, 100079109, 100083472; CD2359-40QC/100078652, 100079156, 100079157, 100079337, 100083407, 100083465, 100117580, 100127092, 100127103; CD2363-40C/100080416; CD2363-40Q/100080407; CD2391-40C/100080784; CD2391-40QC/100080744

Reason for recall

The firm is releasing new firmware for high voltage implantable cardiac devices (ICD) and cardiac resynchronization therapy defibrillators (CRT-D) intended to strengthen security by providing an additional layer of protection against unauthorized device access. For older generation devices not capable of accepting the firmware, the firm is advising customers of the option to permanently disable the RF communication capability in these devices.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2321-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.