VersaTREK Windows Software — Class III medical device recall Z-2323-2012

Terminated recall by Trek Diagnostic Systems, reported 2012-09-12, distributed in AL, AR, AZ, CA, CO, CT, FL, GA…. Review of instrument software databases and investigation of two customer reports determined that on rare occa

Recall numberZ-2323-2012
Product typeMedical device
ClassificationClass III
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmTrek Diagnostic Systems, Cleveland, OH, United States
Recall initiated2012-07-23
Classified2012-09-06
Reported by FDA2012-09-12
Terminated2013-06-18
DistributedAL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, NC, NE, NJ, NY, OH, OK, OR, PA, PR, RI, SD, TN, TX, VA, VT, WA, WI, WV, WY
Countries outside the USBD, BR, BS, CA, CL, CN, CO, CZ, ES, FR, GB, IE, IN, IT, JO, JP, KE, KW, MX, MY, PE, PH, PK, PT, TH, TR, ZA
Quantity333 units
Reason classessoftware
Model numbers6133-30-3

Product

VersaTREK Windows Software, Version 5.4.3 is a component of the VersaTrek System. The Software CD will be individually packaged using a standard CD plastic case and placed into a standard CD mailer The VersaTREK System is for cultivating and recovering microorganisms, especially bacteria and yeasts, from blood and other normally sterile body fluids.

Reason for recall

Review of instrument software databases and investigation of two customer reports determined that on rare occasions (less than 0.3%), following the removal of a positive signaling bottle from the VersaTREK instrument, the positive test result inadvertently reverted to a negative result in the instrument database.