Data Foundry

Medical device recalls 2025

ENDOTAK RELIANCE G/SG with 4-SITE Connector — Class I medical device recall Z-2326-2025

Ongoing recall by Boston Scientific Corporation, reported 2025-08-27, distributed nationwide. Affected lead models may experience gradually rising LVSI measurements in association with calcification of th

Recall numberZ-2326-2025
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmBoston Scientific Corporation, Saint Paul, MN, United States
Recall initiated2025-07-24
Classified2025-08-20
Reported by FDA2025-08-27
DistributedNationwide (US)
UPC / GTIN / UDI-DI00802526431371, 00802526431388, 00802526431401, 00802526431418, 00802526431432, 00802526431449, 00802526431463, 00802526431470, 00802526474668, 00802526474699, 00802526474705, 00802526480096, 00802526480164, 00802526480188, 00802526485763, 00802526504181, 00802526504396, 00802526531248, 00802526531255, 00802526531262, 00802526531279, 00802526541742, 00802526541827, 00802526541902, 00802526541988 and 8 more
Model numbers0282, 0283, 0285, 0286, 0292, 0293, 0295, 0296

Product

ENDOTAK RELIANCE G/SG with 4-SITE Connector, transvenous defibrillation lead, Permanent Defibrillator Electrodes; Model Nos. 0282, 0283, 0285, 0286, 0292, 0293, 0295, 0296

Reason for recall

Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2326-2025. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.