Data Foundry

Medical device recalls 2024

MNA-420 ENC HDMI — Class II medical device recall Z-2332-2024

Completed recall by Barco N.V., reported 2024-07-17, distributed in FL. We have identified a potential issue with the SFP+ module supplied by Molex (Part Number: B620050) when connec

Recall numberZ-2332-2024
Product typeMedical device
ClassificationClass II
StatusCompleted
Voluntary or mandatedVoluntary
Recalling firmBarco N.V., Kortrijk, N/A, Belgium
Recall initiated2024-04-24
Classified2024-07-11
Reported by FDA2024-07-17
DistributedFL
Quantity88 units
Serial numbers2530985205, 2530986280, 2530986281, 2530986282, 2530986283, 2530986284, 2530986285, 2530986286, 2530986287, 2530986288, 2530986289, 2530986290, 2530986291, 2530986292, 2530986293, 2530986294, 2530986295, 2530986296, 2530986297, 2530986298, 2530986299, 2530986300, 2530986301, 2530986302, 2530986303 and 63 more

Product

MNA-420 ENC HDMI, Material Number K9303311; cathode ray tube display

Reason for recall

We have identified a potential issue with the SFP+ module supplied by Molex (Part Number: B620050) when connected to the ETH0 slot of either the MNA-6x0 ENC HDMI or MNA-4x0 ENC HDMI. In some instances, this connection may render the first HDMI channel inaccessible, making both the channel and the respective MNA ENC unit unusable. Additionally, readings from the module become unavailable when this issue arises. Similarly, when this Molex module is connected to the ETH0 slot of an MNA-620 ENC ANA, readings may also become inaccessible.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2332-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.