Data Foundry

Medical device recalls 2018

Merlin PCS Programmer Software: 3330 The Merlin PCS programmer — Class II medical device recall Z-2335-2018

Terminated recall by St Jude Medical Inc., reported 2018-07-11, distributed nationwide. The firm is advising physicians that exposure to sub-freezing temperatures during the supply chain process cau

Recall numberZ-2335-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSt Jude Medical Inc., Sylmar, CA, United States
Recall initiated2018-04-16
Classified2018-06-29
Reported by FDA2018-07-11
Terminated2021-02-04
DistributedNationwide (US)
Quantity0
Model numbers3330, 3650

Product

Merlin PCS Programmer Software: 3330 The Merlin PCS programmer (model 3650) is a portable, external device used to support the implanted device, and for this recall the software that resides on the programmer (software model 3330) is the method of delivery of the new firmware.

Reason for recall

The firm is advising physicians that exposure to sub-freezing temperatures during the supply chain process caused a transient battery voltage drop for a small number of Confirm Rx Model DM3500 Insertable Cardiac Monitoring (ICM) devices. This drop is an expected and normal behavior for this battery chemistry, but causes an incorrect display of a low battery indicator even after the battery voltage returns to normal.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2335-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.