Data Foundry

Medical device recalls 2024

The VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System is… — Class II medical device recall Z-2338-2024

Ongoing recall by Maquet Cardiovascular, LLC, reported 2024-07-17. There were 27 complaints between March 22, 2024, and April 30, 2024, related to VH-3500 and VH-4000, reporting

Recall numberZ-2338-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmMaquet Cardiovascular, LLC, Wayne, NJ, United States
Recall initiated2024-05-31
Classified2024-07-11
Reported by FDA2024-07-17
UPC / GTIN / UDI-DI00607567700406, 00607567700451
Lot numbers3000362934, 3000371260, 3000371798, 3000372477, 3000376542, 3000377320, 3000378288, 3000378554, 3000378950, 3000380314, 3000380847

Product

The VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System is designed for use in conjunction with the 7 mm Extended Length Endoscope. The Harvesting Cannula has four lumens which house the Endoscope, C-Ring, distal lens washer tube, and harvesting tool for cutting and cauterizing vessel branches. The harvesting tool has two curved jaws. One of the jaws contains the heating elements for branch cutting and cauterization and spot cautery. An area near the tip of the convex side of the jaw is used for spot cautery. The activation toggle is used to control the jaws to activate the heating elements. Positioning of the device, cutting, and cauterization are performed under endoscopic visualization.

Reason for recall

There were 27 complaints between March 22, 2024, and April 30, 2024, related to VH-3500 and VH-4000, reporting that the C-Ring wire was straightened (due to insufficient bend radius) and that the distance between C-Ring and Harvesting Tool was closer than normal. No adverse events have been reported to date in direct association with this issue.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2338-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.