Tracheal Intubation Fiberscope OLYMPUS LF-P- To be used with an… — Class II medical device recall Z-2341-2023
Ongoing recall by Olympus Corporation of the Americas, reported 2023-08-16, distributed nationwide. Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatibl
| Recall number | Z-2341-2023 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Olympus Corporation of the Americas, Center Valley, PA, United States |
| Recall initiated | 2023-05-26 |
| Classified | 2023-08-04 |
| Reported by FDA | 2023-08-16 |
| Distributed | Nationwide (US) |
| Quantity | 686 units |
| Reason classes | labeling |
| UPC / GTIN / UDI-DI | 04953170051098 |
Product
Tracheal Intubation Fiberscope OLYMPUS LF-P- To be used with an Olympus Light Source, documentation equipment, display monitor and other ancillary equipment for management of Endotracheal Tube
Reason for recall
Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights the following changes: 1. Removal of LF-V and LF-P as compatible with the OER 2. Change conditions for ETO gas sterilization from 12% EtO to 100% EtO 3. Removing the reusable cleaning brush for the manual cleaning. Single Use brushes remain compatible