Data Foundry

Medical device recalls 2023

Tracheal Intubation Fiberscope OLYMPUS LF-P- To be used with an… — Class II medical device recall Z-2341-2023

Ongoing recall by Olympus Corporation of the Americas, reported 2023-08-16, distributed nationwide. Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatibl

Recall numberZ-2341-2023
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmOlympus Corporation of the Americas, Center Valley, PA, United States
Recall initiated2023-05-26
Classified2023-08-04
Reported by FDA2023-08-16
DistributedNationwide (US)
Quantity686 units
Reason classeslabeling
UPC / GTIN / UDI-DI04953170051098

Product

Tracheal Intubation Fiberscope OLYMPUS LF-P- To be used with an Olympus Light Source, documentation equipment, display monitor and other ancillary equipment for management of Endotracheal Tube

Reason for recall

Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights the following changes: 1. Removal of LF-V and LF-P as compatible with the OER 2. Change conditions for ETO gas sterilization from 12% EtO to 100% EtO 3. Removing the reusable cleaning brush for the manual cleaning. Single Use brushes remain compatible

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2341-2023. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.