Data Foundry

Medical device recalls 2024

191014: 2008K@HOME HD SYSTEM W/BIBAG — Class II medical device recall Z-2342-2024

Ongoing recall by Fresenius Medical Care Holdings, Inc., reported 2024-07-17, distributed nationwide. An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period

Recall numberZ-2342-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmFresenius Medical Care Holdings, Inc., Waltham, MA, United States
Recall initiated2024-06-03
Classified2024-07-11
Reported by FDA2024-07-17
DistributedNationwide (US)
Countries outside the USCA
Quantity20 units
Serial numbers6K0S162351, 6K0S162352, 6K0S162353, 6K0S162354, 6K0S162355, 6K0S169113, 6K0S169114, 6K0S169115, 6K0S169116, 6K0S169117, 6K0S169118, 6K0S169119, 6K0S169120, 6K0S169121, 6K0S169122, 6K0S169123, 6K0S169124, 6K0S169125, 6K0S169126, 6K0S169127
Model numbers191014, F40015481

Product

191014: 2008K@HOME HD SYSTEM W/BIBAG (CANADA) - Blood Pump Rotor spare part, Model Number: F40015481 Rev A

Reason for recall

An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2342-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.