Data Foundry

Medical device recalls 2012

Interfaces subject to recall: Medtronic NIM — Class II medical device recall Z-2343-2012

Terminated recall by Medtronic Xomed, Inc., reported 2012-09-19, distributed nationwide. In January 2010, Medtronic issued a recall of Medtronic NIM 3.0 Patient Interfaces after Medtronic identified

Recall numberZ-2343-2012
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmMedtronic Xomed, Inc., Jacksonville, FL, United States
Recall initiated2010-09-27
Classified2012-09-10
Reported by FDA2012-09-19
Terminated2013-11-26
DistributedNationwide (US)
Countries outside the USAE, AT, AU, BE, CA, CZ, DE, DK, ES, FI, FR, GB, GR, GT, HK, HR, IE, IL, IN, IR, IT, JP, LB, LK, LU, NL, NO, NZ, PT, SE, SG, TR
Quantity53
Model numbers3.0, 32MA, 8253210, 8253410

Product

Interfaces subject to recall: Medtronic NIM (Nerve Integrity Monitor) 3.0 Patient Interfaces REF 8253410 (Neuro). LABELING: Box Label: REF 8253410 x1***NIM-Response 3.0 Patient Interface***Medtronic Xomed, Inc., 6743 Southpoint Drive North Jacksonville, FL 32216-0980 USA***Rx Only***Lot***SN:***" Top of unit is labeled in part: "***NIM-Response 3.0***Medtronic***" Bottom/Underside of unit is labeled in part: "***5X20mm FA.L., 32mA. 250VAC***REF***MEDTRONIC XOMED INC. JACKSONVILLE, FL USA***" Cord is labeled in part: "***Ref.***Mfg Date***Sn***" The NIM 3.0 Patient Interfaces [REF #'s 8253210 (Pulse), 8253200 (Response), 8253410 (Neuro)] is an EMG Nerve Integrity Monitor intended for locating and identifying cranial and peripheral motor and mixed motor-sensory nerves during surgery, including spinal cord and spinal nerve roots.

Reason for recall

In January 2010, Medtronic issued a recall of Medtronic NIM 3.0 Patient Interfaces after Medtronic identified a trend of open channel conditions (signal loss from the electrode). When the NIM 3.0 detects the signal loss from the electrode (i.e. open channel), it fails the electrode check and provides an audible and visual warning to the user which could also create a delay in the surgical procedu

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2343-2012. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.