RayStation 8B — Class II medical device recall Z-2346-2020
Terminated recall by RAYSEARCH LABORATORIES AB, reported 2020-06-17, distributed nationwide. Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algori
| Recall number | Z-2346-2020 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | RAYSEARCH LABORATORIES AB, Stockholm, N/A, Sweden |
| Recall initiated | 2020-05-18 |
| Classified | 2020-06-10 |
| Reported by FDA | 2020-06-17 |
| Terminated | 2021-08-06 |
| Distributed | Nationwide (US) |
| Reason classes | software |
| UPC / GTIN / UDI-DI | 07350002010129 |
Product
RayStation 8B, UDI # 07350002010129 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.
Reason for recall
Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2346-2020. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.