Data Foundry

Medical device recalls 2016

Richard Allan Scientific 10% Neutral Buffered Formalin — Class II medical device recall Z-2351-2016

Terminated recall by Richard-Allan Scientific Company, reported 2016-08-17, distributed in AZ, CA, FL, MD, NY, PA. potential contamination with potassium hydroxide may have an impact when performing immunohistochemistry testi

Recall numberZ-2351-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmRichard-Allan Scientific Company, Kalamazoo, MI, United States
Recall initiated2016-06-13
Classified2016-08-05
Reported by FDA2016-08-17
Terminated2016-11-08
DistributedAZ, CA, FL, MD, NY, PA
Quantity18 cases
Lot numbers371348, 371355
Model numbers53901, 561201
Expiration dates2020-04-01

Product

Richard Allan Scientific 10% Neutral Buffered Formalin, Part Number: 53901 and 561201 Item 53901 90ml bottle, 125 per case; Item 561201 120ml bottle, 75 per case.

Reason for recall

potential contamination with potassium hydroxide may have an impact when performing immunohistochemistry testing. It is unknown whether the contamination will affect tissue antigenicity. If it were to affect tissue antigenicity, the outcome would be risk of false positive or false negative test results which may lead to serious injury. There is the possibility the patient may need to be re-biopsied.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2351-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.