Data Foundry

Medical device recalls 2024

BD Multitest 6-Color TBNK CE-IVD — Class II medical device recall Z-2351-2024

Ongoing recall by Becton, Dickinson and Company, BD Biosciences, reported 2024-07-24, distributed in AL, AR, CA, CT, DC, DE, FL, GA…. Multitest with compromised raw material used to assess immune system status may result in flow cytometry plot

Recall numberZ-2351-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmBecton, Dickinson and Company, BD Biosciences, San Jose, CA, United States
Recall initiated2024-05-08
Classified2024-07-12
Reported by FDA2024-07-24
DistributedAL, AR, CA, CT, DC, DE, FL, GA, ID, IL, IN, KY, LA, MA, MD, MI, MN, NC, NE, NH, NJ, NM, NV, NY, OH, OR, PA, RI, SD, TN, TX, UT, WA, WI
Quantity10,027
UPC / GTIN / UDI-DI00382903371662, 00382906446114, 00382906629678, 00382906629951
Lot numbers17582, 22709, 26797, 32674, 337166/382903371662/17569, 49506, 49510, 49517, 644611/382906446114/20010, 662967/382906629678/17572, 662995/382906629951/49508, 69761
Model numbers337166, 6-COLOR, 644611, 662967, 662995

Product

BD Multitest 6-Color TBNK CE-IVD, REF: 644611; BD Multitest 6-Color TBNK Kit with BD Trucount Tubes CE-IVD, REF: 337166; BD Multitest 6-Color TBNK, IVD, REF: 662967; BD Multitest 6-Color TBNK Kit with BD Trucount Tubes IVD, REF: 662995. BD Multitest" 6-Color TBNK reagent with optional BD Trucount" Tubes is a six-color direct immunofluorescence reagent.

Reason for recall

Multitest with compromised raw material used to assess immune system status may result in flow cytometry plot with unusual staining pattern, nonspecific aggregates, high background, double positive event increase on CD3+CD4+vsCD3+CD8+ and/or CD3-CD16+CD56+vsCD3-CD19+, contaminated antibodies, may lead to erroneous or delayed results, additional patient sample, might impact healthcare decisions