Data Foundry

Medical device recalls 2019

ACUSON NX3 Diagnostic Ultrasound System — Class II medical device recall Z-2352-2019

Ongoing recall by Siemens Medical Solutions USA, Inc., reported 2019-08-28, distributed nationwide. The action is being initiated due to internal testing which identified a possibility for transducers to exceed

Recall numberZ-2352-2019
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmSiemens Medical Solutions USA, Inc., Mountain View, CA, United States
Recall initiated2019-01-28
Classified2019-08-22
Reported by FDA2019-08-28
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BD, BE, BG, BH, BO, BR, CA, CH, CL, CN, CO, CR, CZ, DE, DK, DO, DZ, EC, EG, ES, FI, FR, GB, GR, GT, HK, HR, HU, ID, IE, IL, IN, IQ, IR, IT, JO, JP, KE, KR, LB, LK, LU, LV, MA, MX, MY, NL, NO, NP, PA, PH, PK, PL, PT, RO, RS, SA, SE, SG, SI, SK, SV, TH, TR, TW, UA, UY, VN, ZA
Quantity1,384 systems
UPC / GTIN / UDI-DI04056869004297
Model numbers11235612

Product

ACUSON NX3 Diagnostic Ultrasound System, Model # 11235612, RX Only, Siemens Medical Solutions, containing software versions: VA10A, VA10B, VA10C, VA10D, VA10E, VA10F, VB20A, and VB20B; UDI # 04056869004297 The ACUSON NX2 diagnostic ultrasound imaging systems are intended for the following applications: Cardiac, Fetal, Abdominal (including liver), Pediatric, Small Parts (Small Organ), Transcranial, OB/GYN, Pelvic, Urology, Vascular (including Peripheral Vessel) and Musculoskeletal applications. The systems also provide for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.

Reason for recall

The action is being initiated due to internal testing which identified a possibility for transducers to exceed the acoustic output power (AOP) values defined . The transit voltage values based on PW Doppler are calculated normally but not loaded correctly into hardware