Data Foundry

Medical device recalls 2026

SD LTM 64 PLUS — Class II medical device recall Z-2353-2026

Ongoing recall by Micromed S.p.A., reported 2026-06-17, distributed nationwide. Natus has become aware of two complaints related to the SD LTM 64 PLUS where the EEG traces acquired from two

Recall numberZ-2353-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmMicromed S.p.A., Mogliano Veneto, N/A, Italy
Recall initiated2026-05-04
Classified2026-06-05
Reported by FDA2026-06-17
DistributedNationwide (US)
Quantity207 units
Reason classessoftware
UPC / GTIN / UDI-DI08033928120897
Serial numbersBAA-0051/01-19, BAA-0069/02-19, BAA-0070/02-19, BAA-0071/02-19, BAA-0082/02-19, BAA-0083/02-19, BAA-0084/02-19, BAA-0085/02-19, BAA-0086/02-19, BAA-0087/02-19, BAA-0088/02-19, BAA-0089/02-19, BAA-0090/02-19, BAA-0091/02-19, BAA-0092/02-19, BAA-0093/02-19, BAA-0095/02-19, BAA-0096/02-19, BAA-0098/02-19, BAA-0104/02-19, BAA-0105/02-19, BAA-0106/02-19, BAA-0107/02-19, BAA-0108/02-19, BAA-0109/02-19 and 23 more

Product

SD LTM 64 PLUS, EEG Amplifier/recorder, Model/Catalog Number: SD LTM 64 PLUS, Software Version: firmware 2021.02, 2022.01 or 2022.02

Reason for recall

Natus has become aware of two complaints related to the SD LTM 64 PLUS where the EEG traces acquired from two or four different amplifiers used in a multiple amplifier configuration (128 or 256 channels) running firmware version 2022.02 were displayed with a shift of 1 second between the 64channels blocks. In rare circumstances, a fixed 1-second temporal delay may appear between two 64-channel modules with firmware 2021.02, 2022.01 or 2022.02 in a multi-module SEEG configuration, potentially leading to misinterpretation of results.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2353-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.