Data Foundry

Medical device recalls 2016

Alere INRatio PT/INR Test Strips Model Number: 0100071 — Class I medical device recall Z-2361-2016

Terminated recall by Alere San Diego, Inc., reported 2016-08-31, distributed nationwide. Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring Systems (T

Recall numberZ-2361-2016
Product typeMedical device
ClassificationClass I
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmAlere San Diego, Inc., San Diego, CA, United States
Recall initiated2016-07-11
Classified2016-08-19
Reported by FDA2016-08-31
Terminated2017-07-20
DistributedNationwide (US)
Countries outside the USAR, AU, BD, BE, BR, CA, CL, CN, CO, EG, ES, FR, HK, ID, IL, IN, IT, JP, KR, MY, NL, RU, SG, TH, TT, TW, UA, UY, VN, ZA
Model numbers0100071, 0100139

Product

Alere INRatio PT/INR Test Strips Model Number: 0100071, 0100139 Product Usage: The INRatio/INRatio2 Test Strips perform a modified version of the one-stage Prothrombin Time test, using a recombinant human thromboplastin reagent. The clot formed in the Prothrombin Time reaction is detected by a change in the electrical impedance of the sample during the coagulation process. The system consists of a monitor and disposable test strips. The monitor measures impedance, heats the test strip to the proper reaction temperature, and provides a user interface. The blood sample is applied to the Test Strip and the clotting reaction occurs on the Test Strip.

Reason for recall

Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2361-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.