Brand Name: EXENT Analyser Product Name: EXENT Analyser Model/Catalog… — Class II medical device recall Z-2362-2026
Ongoing recall by The Binding Site Group, Ltd., reported 2026-06-17, distributed nationwide. It was reported that plates 2 and 3 were marked as passed , although it appeared that QC had not been performe
| Recall number | Z-2362-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | The Binding Site Group, Ltd., Birmingham, N/A, United Kingdom |
| Recall initiated | 2026-04-16 |
| Classified | 2026-06-09 |
| Reported by FDA | 2026-06-17 |
| Distributed | Nationwide (US) |
| Quantity | 1 |
| UPC / GTIN / UDI-DI | 05051700020886 |
| Serial numbers | 101150 |
| Model numbers | IE800.A, IE800A |
Product
Brand Name: EXENT Analyser Product Name: EXENT Analyser Model/Catalog Number: IE800.A Software Version: 1.0.20 Product Description: The EXENT Analyser is an automated analyser intended for the qualitative and quantitative in vitro measurement of analytes in human body fluids used in conjunction with the EXENT assays.
Reason for recall
It was reported that plates 2 and 3 were marked as passed , although it appeared that QC had not been performed on these plates, despite samples being measured. The potential clinical hazards include false or missed detection of M-protein (false positive); and falsely elevated or decreased M-protein concentration.