Alinity ci -series System Control¿Module — Class II medical device recall Z-2363-2019

Terminated recall by Abbott Gmbh & Co. KG, reported 2019-09-04, distributed nationwide. Abbott has identified an issue with all on market versions (v2.6.0 and v2.6.1) of Alinity ci series Software w

Recall numberZ-2363-2019
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmAbbott Gmbh & Co. KG, Wiesbaden, N/A, Germany
Recall initiated2019-06-10
Classified2019-08-23
Reported by FDA2019-09-04
Terminated2023-06-14
DistributedNationwide (US)
Quantity1,763 units
Reason classessoftware
Model numbers03R70

Product

Alinity ci -series System Control¿Module, Part Number 03R70 01 Product Usage: The Alinity ci series System Control Module has a scalable design to provide full integration of clinical chemistry and immunoassay analysis. The Alinity ci series System Control Module includes a clinical chemistry module and an immunoassay module, each performing all sample processing activities, and a system control module to provide a single user friendly interface. Each system can be customized by combining one or more sample processing modules, in multiple configurations, with a system control module to form a single workstation. The Alinity i Processing Module is a fully automated immunoassay analyzer allowing random and continuous access, as well as priority and automated retest processing using chemiluminescent microparticle immunoassay (CMIA) technology. CMIA technology is used to determine the presence of antigens, antibodies, and analytes in samples.

Reason for recall

Abbott has identified an issue with all on market versions (v2.6.0 and v2.6.1) of Alinity ci series Software where incorrect results may occur after a system Stop due to the Alinity i re use of reaction vessels (RVs). This issue only occurs if the system is transitioned from Processing to Stopped to Idle.