Alinity ci -series System Control¿Module — Class II medical device recall Z-2363-2019
Terminated recall by Abbott Gmbh & Co. KG, reported 2019-09-04, distributed nationwide. Abbott has identified an issue with all on market versions (v2.6.0 and v2.6.1) of Alinity ci series Software w
| Recall number | Z-2363-2019 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Abbott Gmbh & Co. KG, Wiesbaden, N/A, Germany |
| Recall initiated | 2019-06-10 |
| Classified | 2019-08-23 |
| Reported by FDA | 2019-09-04 |
| Terminated | 2023-06-14 |
| Distributed | Nationwide (US) |
| Quantity | 1,763 units |
| Reason classes | software |
| Model numbers | 03R70 |
Product
Alinity ci -series System Control¿Module, Part Number 03R70 01 Product Usage: The Alinity ci series System Control Module has a scalable design to provide full integration of clinical chemistry and immunoassay analysis. The Alinity ci series System Control Module includes a clinical chemistry module and an immunoassay module, each performing all sample processing activities, and a system control module to provide a single user friendly interface. Each system can be customized by combining one or more sample processing modules, in multiple configurations, with a system control module to form a single workstation. The Alinity i Processing Module is a fully automated immunoassay analyzer allowing random and continuous access, as well as priority and automated retest processing using chemiluminescent microparticle immunoassay (CMIA) technology. CMIA technology is used to determine the presence of antigens, antibodies, and analytes in samples.
Reason for recall
Abbott has identified an issue with all on market versions (v2.6.0 and v2.6.1) of Alinity ci series Software where incorrect results may occur after a system Stop due to the Alinity i re use of reaction vessels (RVs). This issue only occurs if the system is transitioned from Processing to Stopped to Idle.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2363-2019. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.