Data Foundry

Medical device recalls 2020

VITROS Chemistry Product Performance Verifier I — Class II medical device recall Z-2366-2020

Terminated recall by Ortho-Clinical Diagnostics, reported 2020-06-24, distributed nationwide. Multiple coatings of VITROS Chemistry Products AMYL Slides are showing imprecision on control fluids when comp

Recall numberZ-2366-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmOrtho-Clinical Diagnostics, Rochester, NY, United States
Recall initiated2020-05-11
Classified2020-06-15
Reported by FDA2020-06-24
Terminated2021-09-22
DistributedNationwide (US)
Countries outside the USMX
Reason classespotency
UPC / GTIN / UDI-DI10758750004577

Product

VITROS Chemistry Product Performance Verifier I, Product Code 8231474, UDI # 10758750004577 - Product Usage: For in vitro diagnostic use only to quantitatively measure amylase activity in serum, plasma, and urine using VITROS 250/350/5, 1 FS/4600/XT 3400 Chemistry Systems and VITROS 5600/XT 7600 Integrated Systems.

Reason for recall

Multiple coatings of VITROS Chemistry Products AMYL Slides are showing imprecision on control fluids when compared to the expected within lab SD as listed on the Performance Verifier (PV) Assay Sheets.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2366-2020. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.