Data Foundry

Medical device recalls 2012

Sterile Distraction Screws of 3 different sizes: 12mm — Class II medical device recall Z-2385-2012

Terminated recall by Tedan Surgical Innovations Llc, reported 2012-09-26, distributed nationwide. This letter is to notify you that several of our sterile screw products are being voluntarily recalled. Produc

Recall numberZ-2385-2012
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmTedan Surgical Innovations Llc, Sugar Land, TX, United States
Recall initiated2012-06-11
Classified2012-09-14
Reported by FDA2012-09-26
Terminated2013-07-31
DistributedNationwide (US)
Countries outside the USAU, TW
Reason classessterility
Lot numbers019715, 019716, 019942, 019943, 019977, 019978, 020102, 020103, 020104, 020105, 020106, 020107, 020108, 020609, 020782, 021539, 021663, 022097, 022098, 022099, 022122, 022123, 022124, 022371, 022372, 022790, 022791, 022792, 022794, 022795, 022796, 022827, 022828, 023266

Product

Sterile Distraction Screws of 3 different sizes: 12mm, 14mm and 16mm sold under TSI, 3COR, DB Surgical, and American Surgical Products brands and Sterile XL Screws of 7 different sizes: 90mm, 100mm, 110mm, 120mm, 130mm, 140mm, 150mm sold under TSI brand Sterile Distraction Screws are intended for the distraction in the anterior approach to the cervical spine. Sterile XL Screws are intended for the fixation of the blade to the vertebral body to prohibit drifting in the extreme lateral approach to the lumbar spine.

Reason for recall

This letter is to notify you that several of our sterile screw products are being voluntarily recalled. Product sterility may be compromised due to a mechanical failure of the protective packaging tube. Tube end plugs may dislodge creating the potential for the screw to come out of the protective tube and then the sharp end of the screw pierce the outer sterile barrier pouch.