Data Foundry

Medical device recalls 2021

DIAMONDBACK 360 GEN2 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM — Class II medical device recall Z-2385-2021

Terminated recall by Cardiovascular Systems Inc, reported 2021-09-08, distributed nationwide. There is a potential mislabeling of the OAD crown sizes: 1.50mm Sold Crown OADs may be packaged s 1.25mm Micro

Recall numberZ-2385-2021
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmCardiovascular Systems Inc, Saint Paul, MN, United States
Recall initiated2021-07-21
Classified2021-08-28
Reported by FDA2021-09-08
Terminated2026-03-18
DistributedNationwide (US)
Quantity8 units
Reason classeslabeling
Lot numbers340925
Model numbersDBP-150SOLID145

Product

DIAMONDBACK 360 GEN2 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM, 1.5MM SOLID, 145 CM, REF DBP-150SOLID145

Reason for recall

There is a potential mislabeling of the OAD crown sizes: 1.50mm Sold Crown OADs may be packaged s 1.25mm Micro Crown OADs and vice versa.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2385-2021. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.