Elecsys Vitamin D total II — Class II medical device recall Z-2389-2018
Terminated recall by Roche Diagnostics Corporation, reported 2018-07-18, distributed nationwide. The device may give a falsely elevated result that is non-reproducible. If vitamin D level is monitored by lab
| Recall number | Z-2389-2018 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Roche Diagnostics Corporation, Indianapolis, IN, United States |
| Recall initiated | 2018-04-27 |
| Classified | 2018-07-08 |
| Reported by FDA | 2018-07-18 |
| Terminated | 2020-05-11 |
| Distributed | Nationwide (US) |
| Quantity | 347 units |
Product
Elecsys Vitamin D total II, Vitamin D Test System, Material number 07028148190 Product Usage: This assay is intended for the quantitative determination of total 25-hydroxyvitamin D in human serum and plasma. This assay is to be used as an aid in the assessment of vitamin D sufficiency in adults. The electrochemiluminescence binding assay is intended for use on the immunoassay analyzer.
Reason for recall
The device may give a falsely elevated result that is non-reproducible. If vitamin D level is monitored by laboratory tests and adapted based on the result, a dosage reduction of vitamin D supplementation cannot be excluded, and as a consequence, vitamin D deficiency may develop over time.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2389-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.