Data Foundry

Medical device recalls 2018

Elecsys Vitamin D total II — Class II medical device recall Z-2390-2018

Terminated recall by Roche Diagnostics Corporation, reported 2018-07-18, distributed nationwide. The device may give a falsely elevated result that is non-reproducible. If vitamin D level is monitored by lab

Recall numberZ-2390-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmRoche Diagnostics Corporation, Indianapolis, IN, United States
Recall initiated2018-04-27
Classified2018-07-08
Reported by FDA2018-07-18
Terminated2020-05-11
DistributedNationwide (US)
Quantity3,591 units

Product

Elecsys Vitamin D total II, Vitamin D Test System, Material number 07464215160 Product Usage: This assay is intended for the quantitative determination of total 25-hydroxyvitamin D in human serum and plasma. This assay is to be used as an aid in the assessment of vitamin D sufficiency in adults. The electrochemiluminescence binding assay is intended for use on the immunoassay analyzer.

Reason for recall

The device may give a falsely elevated result that is non-reproducible. If vitamin D level is monitored by laboratory tests and adapted based on the result, a dosage reduction of vitamin D supplementation cannot be excluded, and as a consequence, vitamin D deficiency may develop over time.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2390-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.