Reliance 1227 Cart and Utensil Washer/Disinfector — Class II medical device recall Z-2392-2018

Terminated recall by Steris Corporation, reported 2018-07-18, distributed nationwide. The firm has become aware that the Reliance 1227 Cart and Utensil Washer/Disinfector s Chemical Low Level alar

Recall numberZ-2392-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSteris Corporation, Mentor, OH, United States
Recall initiated2018-05-23
Classified2018-07-09
Reported by FDA2018-07-18
Terminated2019-11-05
DistributedNationwide (US)
Countries outside the USCA
Quantity118
Reason classeschemical contamination

Product

Reliance 1227 Cart and Utensil Washer/Disinfector, FW03101, FW03102, FW03S003 Product Usage: The Reliance 1227 Cart and Utensil Washer/Disinfector is intended for use in the cleaning and low-level disinfection of bedpans and urinals, basins, carts, beds, theatre shoes and other miscellaneous reusable items used in the care of patients.

Reason for recall

The firm has become aware that the Reliance 1227 Cart and Utensil Washer/Disinfector s Chemical Low Level alarm, intended to prevent the user from initiating a cycle when a low chemical level situation occurs, does not operate as intended. Currently, if a low chemical level situation occurs, the alarm will only be generated at the unit s next power-up or when accessing service mode. The failure of the Reliance 1227 to identify that the chemicals used for cleaning are low or empty could result in bedpans and urinals, basins, case carts, beds, theater shoes and other miscellaneous reusable patient care items not being properly cleaned before disinfection or reuse.