Data Foundry

Medical device recalls 2018

Femoral Nail — Class II medical device recall Z-2396-2018

Terminated recall by Stryker GmbH, reported 2018-07-18, distributed in FL, KY, NC, TN, UT, WA. The manufacturer has discovered that potentially out-of-specification products may have left the factory. Non-

Recall numberZ-2396-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmStryker GmbH, Selzach, Switzerland
Recall initiated2018-04-27
Classified2018-07-10
Reported by FDA2018-07-18
Terminated2020-05-21
DistributedFL, KY, NC, TN, UT, WA
Quantity9
Reason classespotency, device malfunction
UPC / GTIN / UDI-DI04546540199775
Lot numbersK081650

Product

Femoral Nail, A/R T2 Femur ¿12x360 mm, Product Number 18251236S

Reason for recall

The manufacturer has discovered that potentially out-of-specification products may have left the factory. Non-conforming cannulation of the nails may result in reduced component strength and potentially premature nail breakage.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2396-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.