Infinity FMS Pak — Class II medical device recall Z-2408-2023
Ongoing recall by Alcon Research, LTD., reported 2023-08-23, distributed nationwide. Reports of patient injury during use in procedures due to potential for plastic debris from over-tightening wi
| Recall number | Z-2408-2023 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Alcon Research, LTD., Sinking Spring, PA, United States |
| Recall initiated | 2023-06-28 |
| Classified | 2023-08-16 |
| Reported by FDA | 2023-08-23 |
| Distributed | Nationwide (US) |
| Countries outside the US | AR, AT, AU, AW, BB, BD, BE, BG, BH, BM, BR, BS, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DZ, EC, EG, ES, FI, FR, GB, GH, GR, HK, HN, HR, HU, ID, IE, IL, IN, IQ, IR, IS, IT, JM, JO, JP, KE, KR, KW, KY, LB, LK, LU, LV, MA, MT, MX, MY, NI, NL, NO, NZ, OM, PA, PE, PH, PK, PL, PT, QA, RO, RS, SA, SE, SG, SI, SK, SV, TH, TN, TR, TT, TW, UA, UY, VN, ZA |
| Quantity | 51,054 units |
| UPC / GTIN / UDI-DI | 00380657410859, 00380657502806, 00380657502813, 00380657520862, 00380657520879, 00380657520909, 00380657520930, 00380657520947, 00380657520954, 00380657520961 |
| Lot numbers | 14VCLK, 14VCLL, 14VCLM, 14WCPD, 14XPL9, 14XPLA, 14Y62U, 14Y62V, 14YLLD, 14YLLE, 14YXXD, 150374, 15037P, 15037R, 15037T, 150FEL, 150FEM, 150FEN, 150FEP, 150FER, 150FET, 150FEU, 150FEV, 150FEW, 150FF5, 150FF6, 150FFX, 151J41, 151J42, 151J5W, 151VEY, 152H82, 152W36, 1532WH, 1532XL, 1532XU, 1532Y0, 1532Y5, 1532YC, 1532YE and 77 more |
| Model numbers | 8065741085, 8065750280, 8065750281, 8065752086, 8065752087, 8065752090, 8065752093, 8065752094, 8065752095, 8065752096 |
Product
Infinity FMS Pak
Reason for recall
Reports of patient injury during use in procedures due to potential for plastic debris from over-tightening with included plastic wrench.