Data Foundry

Medical device recalls 2018

FDI FORT DEFIANCE INDUSTRIES P2131 AUTOMATED FIELD STEAM STERILIZER — Class II medical device recall Z-2411-2018

Terminated recall by Fort Defiance Industries, LLC, reported 2018-07-18, distributed nationwide. During long-term storage (i.e. military depot), the interaction of stagnant water and dissimilar metals in the

Recall numberZ-2411-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmFort Defiance Industries, LLC, Loudon, TN, United States
Recall initiated2018-05-24
Classified2018-07-10
Reported by FDA2018-07-18
Terminated2023-06-27
DistributedNationwide (US)
Quantity199 units
Serial numbersAFS-0001
Model numbersP2131

Product

FDI FORT DEFIANCE INDUSTRIES P2131 AUTOMATED FIELD STEAM STERILIZER, NSN: 6530-01-641-4641 Product Usage: The Fort Defiance Industries Automated Field Steam Sterilizer Model P2131 is designed for sterilization of porous and non-porous, heat- and moisture-stable materials (e.g., surgical instruments and textiles) used in healthcare facilities. The P2131 sterilizer is a transportable device designed for use in a variety of austere environments.

Reason for recall

During long-term storage (i.e. military depot), the interaction of stagnant water and dissimilar metals in the plumbing of the P2131 device produces deposits in the water system that can potentially interfere with the rear sight gauge and / or the water liquid level switch.